BioPlus 1270/1400

ATEX-marking
II 3G Ex nA nC nL llB T3 Gc

The largest cabinets in the BioLine range, the BioPlus 1270 and 1400 are large by any stretch of imagination. Exceptional capacity and means to store a vast amount of items, make these models applicable to a very wide variety of uses where storage capacity and performance is essential.

Like all the other BioPlus models, the 1270 and 1400 can cope with anything from general purpose to mission-critical storage scenarios. The BioPlus 1270 and 1400 are available as a refrigerator ER (-2°C/+20°C) or freezer RF (-25°C/-5°C), and comes with stainless steel interior and white or stainless steel exterior.

 

Scroll down to the accessory panel to see what fittings you can equip the cabinet with.

ER1270

Temperature range -2°C/+20°C
Ambient temperature range Solid door 10°C/43°C; glass door 10°C/38°C
Control unit Gram Control Unit with voltage free contact, E-sensor, dry cooling and offset function
Alarms Acoustic and visual, High/Low temperature alarms and door alarm
Access port 1 pcs. Ø 24.5 mm
Material interior Stainless steel
Material exterior White lacquered steel or stainless steel
Dimensions (W x D x H) 1390 x 876 x 1875/2125*
Net weight Unpackaged net weight without optional fittings: 271 kg
Gross volume 1270 litres
Net volume 864 litres
Modules for shelves and drawers Shelves 48 - Drawers 24
Insulation 60 mm polyurethane with HFC-free cyclopentane propellant
K-Value Solid door 0.3125 W/m²K; glass door 0.504 W/m²K
Refrigerant R290/ R134a/prepared for external compressor/prepared for glycol
CO2e R134a: 508 / R134a (Dual comp.) 615
Refrigeration capacity at -10°C R290 946 watt; R134a 703 watt
Energy consumption (kWh/24h) (R290) Solid door: 2.50 / Glass door: 3.48 | Dual comp. solid door: 2.57 / Dual comp. glass door: N/A
Energy consumption (kWh/24h) (R134a) Solid door: 2.85 / Glass door: 3.39 | Dual comp. solid door: N/A / Dual comp. glass door: 5.09
Heat emission (Watt) (R290) Solid door: 530,44 / Glass door: 540,1 | Dual comp. solid door: 746,5 / Dual comp. glass door: N/A**
Heat emission (Watt) (R134a) Solid door: 577,56 / Glass door: 557,75 | Dual comp. solid door: N/A / Dual comp. glass door: 468**
Connection 230 V, 50 Hz
Connection load (Watt) (R290) 596 | Dual comp. 838
Connection load (Watt) (R134a) 580 | Dual comp. 1050
Start current (R290) 13,2 A
Start current (R134a) 14,8 A
Sound level 48.3 dB(A)
Air system BioLine ventilated air distribution system
Defrost system Automatic smart defrost with re-evaporation of defrost water

RF1270

Temperature range -25°C/-5°C
Ambient temperature range 10°C/43°C
Control unit Gram Control Unit with voltage free contact, E-sensors and offset function
Alarms Acoustic and visual, High/Low temperature alarms and door alarm
Access port 1 pcs. Ø 24.5 mm
Material interior Stainless steel
Material exterior White lacquered steel or stainless steel
Dimensions (W x D x H) 1390 x 876 x 1875/2125*
Net weight Unpackaged net weight without optional fittings: 271 kg
Gross volume 1270 litres
Net volume 864 litres
Modules for shelves and drawers Shelves 48 - Drawers 24
Insulation 60 mm polyurethane with HFC-free cyclopentane propellant
K-Value Solid door 0.3125 W/m²K
Refrigerant R290/ prepared for external compressor
CO2e R404A: 1647 / R404A (Dual comp.) 1726
Refrigeration capacity at -25°C R290 950 watt; R404A 947 watt
Energy consumption (kWh/24h) (R290) 7.97 | Dual comp. N/A
Energy consumption (kWh/24h) (R404A) 7.48 | Dual comp. 8.89
Heat emission (Watt) (R290) 828,42 | Dual comp. 828,42**
Heat emission (Watt) (R404A) 778 | Dual comp. 1048**
Connection 230 V, 50 Hz
Connection load (Watt) (R290) 860 | Dual comp. 930
Connection load (Watt) (R404A) 844 | Dual comp. 1118
Start current (R290) 13,2 A
Start current (R404A) 14,8 A
Sound level 51 dB(A)
Air system BioLine ventilated air distribution system
Defrost system Automatic smart defrost with re-evaporation of defrost water

*Dimensions in mm
**100% operation at +23°C (+/- 1K) ambient temperature

ER1400

Temperature range -2°C/+20°C
Ambient temperature range Solid door 10°C/43°C; glass door 10°C/38°C
Control unit Gram Control Unit with voltage free contact, E-sensor, dry cooling and offset function
Alarms Acoustic and visual, High/Low temperature alarms and door alarm
Access port 1 pcs. Ø 24.5 mm
Material interior Stainless steel
Material exterior White lacquered steel or stainless steel
Dimensions (W x D x H) 1390 x 876 x 2025/2275*
Net weight Unpackaged net weight without optional fittings: 283 kg
Gross volume 1400 litres
Net volume 968 litres
Modules for shelves and drawers Shelves 54 - Drawers 26
Insulation 60 mm polyurethane with HFC-free cyclopentane propellant
K-Value Solid door 0.3124 W/m²K; glass door 0.5081 W/m²K
Refrigerant R290/ R134a/prepared for external compressor/prepared for glycol
CO2e R134a: 508 / R134a (Dual comp.) 615
Refrigeration capacity at -10°C R290 946 watt; R134a 703 watt
Energy consumption (kWh/24h) (R290) Solid door: 2.56 / Glass door: 3.72 | Dual comp. solid door: N/A / Dual comp. glass door: 4.47
Energy consumption (kWh/24h) (R134a) Solid door: 2.80 / Glass door: 4.16 | Dual comp. solid door: 2.83 / Dual comp. glass door: 3.90
Heat emission (Watt) (R290) Solid door: 537,63 / Glass door: 539,1 | Dual comp. solid door: N/A / Dual comp. glass door: 769,38**
Heat emission (Watt) (R134a) Solid door: 533,06 / Glass door: 576,86 | Dual comp. solid door: 461,83 / Dual comp. glass door: 451,75**
Connection 230 V, 50 Hz
Connection load (Watt) (R290) 596 | Dual comp. 838
Connection load (Watt) (R134a) 580 | Dual comp. 1050
Start current (R290) 13,2 A
Start current (R134a) 14,8 A
Sound level 47.5 dB(A)
Air system BioLine ventilated air distribution system
Defrost system Automatic smart defrost with re-evaporation of defrost water

RF1400

Temperature range -25°C/-5°C
Ambient temperature range 10°C/43°C
Control unit Gram Control Unit with voltage free contact, E-sensor and offset function
Specification item - Alarms Acoustic and visual, High/Low temperature alarms and door alarm
Access port 1 pcs. Ø 24.5 mm
Material interior Stainless steel
Material exterior White lacquered steel or stainless steel
Dimensions (W x D x H) 1390 x 876 x 2025/2275*
Net weight Unpackaged net weight without optional fittings: 283 kg
Gross volume 1400 litres
Net volume 968 litres
Modules for shelves and drawers Shelves 54 - Drawers 26
Insulation 60 mm polyurethane with HFC-free cyclopentane propellant
K-Value Solid door 0.3124 W/m²K
Refrigerant R290/ R404A/prepared for external compressor
CO2e R404A: 1647 / R404A (Dual comp.) 1726
Refrigeration capacity at -25°C R290 950 watt; R404A 947 watt
Energy consumption (kWh/24h) (R290) 8.39 | Dual comp. N/A
Energy consumption (kWh/24h) (R404A) 8.08 | Dual comp. 9.45
Heat emission (Watt) (R290) 790,36 | Dual comp. 790,36**
Heat emission (Watt) (R404A) 814,5 | Dual comp. 1070,9**
Connection 230 V, 50 Hz
Connection load (Watt) (R290) 860 | Dual comp. 930
Connection load (Watt) (R404A) 844 | Dual comp. 1118
Start current (R290) 13,2 A
Start current (R404A) 14,8 A
Sound level 46.6 dB(A)
Air system BioLine ventilated air distribution system
Defrost system Automatic smart defrost with re-evaporation of defrost water

*Dimensions in mm
**100% operation at +23°C (+/- 1K) ambient temperature

BioPlus 500

All the benefits of the BioPlus range, on a small footprint

BioPlus 500
BioPlus 600/660D

A deep cabinet, with good proportions and performance

BioPlus 600/660D
BioPlus 600/660W

A wide cabinet, enabling superior overview of the items

BioPlus 600/660W
BioPlus 930

Largest single door BioLine cabinet

BioPlus 930
BioPlus EF600W/EF660W

-35°C/-5°C low temperature freezer

BioPlus EF600W/EF660W

Choosing the right product..

“Application” icons bridge the gap between product ranges and the requirements in the different segments, work environments and use patterns. Allowing for quick and precise evaluation of needs and requirements.

CLICK TO EXPLORE APPLICATIONS

Customised to your needs

Every single user has a unique set of requirements for how a biostorage fridge or freezer should be configured, and how the interior should be laid out. That’s why there are no standard arrangements for cabinets in the BioLine range – it’d be highly unlikely that “standard” would live up to your particular working requirements.

Our approach is to make sure that all the BioLine cabinets featured have a high standard specification. 

We then provide you with a comprehensive selection of optional features, as well as standardised modules for internal arrangements, fittings and accessories. 
These choices enable you to select exactly what you need, and to customise both the operating specifications and the interior layout of your BioLine cabinets, so that they’re right in line with your particular requirements.

With a cabinets service life expected to be many years, we argue that using a little more time in selecting the configuration that suites your exact specifications, is worth it.

All BioLine cabinets have a flexible layout, and can be altered on a whim; and if the need should arise, you can easily acquire new fittings according to the changing needs.