BioPlus 500

Application, temperature range, liter and dimensions specifications for BioPlus refrigerators and freezers.

ATEX-marking
II 3G Ex nA nC nL llB T2 Gc

Despite its size, the smallest of the BioPlus models, the BioPlus 500 stands out in its own right. With its width of 60cm, the BioPlus 500 lends itself exceedingly well to situations where performance and reliability are paramount, but wall space is a sparse commodity. Like all other BioPlus models, the 500 can cope with anything from general purpose to mission-critical storage scenarios.

The BioPlus 500 is available as a refrigerator ER (-2°C/+20°C) or freezer RF (-25°C/-5°C), and comes with stainless steel interior, and white or stainless steel exterior. All BioPlus models are fitted with a self-closing door.

Scroll further down to the accessory panel to see what other fittings you can equip the storage cabinet with.

BioPlus premium refrigerator with advanced technology andwhite exterior.
BioPlus extended refrigerator 500 with precise temperature control and an active display.

ER500

Temperature range -2 °C/+20 °C
Ambient temperature range Solid door +10 °C/+43 °C; glass door +10 °C/+38 °C
Control unit Gram Control Unit with voltage free contact, E-sensor, dry cooling and offset function
Alarms Acoustic and visual, High/Low temperature alarms and door alarm
Access port 1 pcs. Ø 24.5 mm
Material interior Stainless steel
Material exterior White lacquered steel or stainless steel
Dimensions (WxDxH) 600 x 805 x 2025/2275*
Net weight Unpackaged net weight without optional fittings: 130 kg
Gross volume 500 litres
Net volume 365 litres
Modules for shelves and drawers Shelves 52 - Drawers 13
Insulation 60 mm polyurethane with HFC-free cyclopentane propellant
K-Value Solid door 0.31 W/m²K; glass door 0.46 W/m²K
Refrigerant R290
Refrigerant charge 95 g
Refrigeration capacity at -10°C R290 389 watt
CO2e -
Energy consumption (kWh/24h) (R290) Solid door: 1.26 / Glass door: 1.67
Heat emission (Watt) (R290) Solid door: 319.8 / Glass door: 315,93**
Connection 230 V, 50 Hz
Connection load (Watt) (R290) 333
Start current (R290) 9,7 A
Sound level 46.2 dB(A)
Air system BioLine ventilated air distribution system
Defrost system Automatic smart defrost with re-evaporation of the defrost water

RF500

Temperature range -25 °C/-5 °C
Ambient temperature range +10 °C/+43 °C
Control unit Gram Control Unit with voltage free contact, E-sensor and offset function
Alarms Acoustic and visual, High/Low temperature alarms and door alarm
Access port 1 pcs. Ø 24.5 mm
Material interior Stainless steel
Material exterior White lacquered steel or stainless steel
Dimensions (WxDxH) 600 x 805 x 2025/2275*
Net weight Unpackaged net weight without optional fittings: 130 kg
Gross volume 500 litres
Net volume 365 litres
Modules for shelves and drawers Shelves 52 - Drawers 13
Insulation 60 mm polyurethane with HFC-free cyclopentane propellant
K-Value Solid door 0.31 W/m²K
Refrigerant R290
Refrigerant charge 92 g
Refrigeration capacity at -25°C R290 374 watt
CO2e -
Energy consumption (kWh/24h) (R290) 3.83
Heat emission (Watt) (R290) 350,69**
Connection 230 V, 50 Hz
Connection load (Watt) (R290) 336
Start current (R290) 10,4 A
Sound level 48.7 dB(A)
Air system BioLine ventilated air distribution system
Defrost system Automatic smart defrost with re-evaporation of defrost water

*     Dimensions in mm
**   100% operation at +23 °C (+/- 1K) ambient temperature

BioPlus 600/660D

A deep cabinet, with good proportions and performance

BioPlus 600/660D
BioPlus 600/660W

A wide cabinet, enabling superior overview of the items

BioPlus 600/660W
BioPlus 930

Largest single door BioLine cabinet

BioPlus 930
BioPlus 1270/1400

Double door models with exceptional capacity

BioPlus 1270/1400
BioPlus EF600W/EF660W

-35°C/-5°C low temperature freezer

BioPlus EF600W/EF660W

Choosing the right product..

“Application” icons bridge the gap between product ranges and the requirements in the different segments, work environments and use patterns. Allowing for quick and precise evaluation of needs and requirements.

Customised to your needs

Every single user has a unique set of requirements for how a biostorage fridge or freezer should be configured, and how the interior should be laid out. That’s why there are no standard arrangements for cabinets in the BioLine range – it’d be highly unlikely that “standard” would live up to your particular working requirements.

Our approach is to make sure that all the BioLine cabinets featured have a high standard specification. 

We then provide you with a comprehensive selection of optional features, as well as standardised modules for internal arrangements, fittings and accessories. 
These choices enable you to select exactly what you need, and to customise both the operating specifications and the interior layout of your BioLine cabinets, so that they’re right in line with your particular requirements.

With a cabinets service life expected to be many years, we argue that using a little more time in selecting the configuration that suites your exact specifications, is worth it.

All BioLine cabinets have a flexible layout, and can be altered on a whim; and if the need should arise, you can easily acquire new fittings according to the changing needs.