BioPlus EF600W/EF660W

ATEX-marking
II 3G Ex nA nC nL llB T2 Gc

Resembling the regular 600W and 660W, the EF is a low temperature cabinet able to run at -35°C. The "EF" (extended freezer) model range comes standard with a handle for secure closing, and PMMA inner doors for maximum cold-retention during door openings. As with the regular ER and RF "W" models, the overview one gets from having a wider cabinet makes for efficient in and output. Like all the other BioPlus models, the EF600W and EF660W models can cope with anything from general purpose to mission-critical storage scenarios. 

The BioPlus EF600W and EF660W are solely available as the -35°C/-5°C model, and comes with stainless steel interior, and white or stainless steel exterior.

 

Scroll down to the accessory panel to see what fittings you can equip the cabinet with.

 

Did you mean the ER or RF model?

EF600W

Temperature range -35 °C/-5 °C
Ambient temperature range +10 °C/+30 °C
Control unit Gram Control Unit with voltage free contact, E-sensor and offset function
Alarms Acoustic and visual, High/Low temperature alarms and door alarm
Access port 1 pcs. Ø 24.5 mm
Material interior Stainless steel and inner doors in PMMA
Material exterior White lacquered steel or stainless steel
Dimensions (W x D x H) 837 x 756 x 1875/2125*
Net weight Unpackaged net weight without optional fittings: 146 kg
Gross volume 600 litres
Net volume 432 litres
Modules for shelves and drawers Shelves 24 - Drawers 12
Insulation 60 mm polyurethane with HFC-free cyclopentane propellant
K-Value Solid door 0.31 W/m²K
Refrigerant R290/prepared for external compressor
Refrigerant charge 92 g
Refrigeration capacity at -40°C R290 244 watt
CO2e -
Energy consumption (kWh/24h) (R290) 8.86 | Water cooled N/A | Dual comp. 12.21
Heat emission (Watt) (R290) 441,69 | Water cooled N/A | Dual comp. 829,5**
Connection 230 V, 50 Hz
Connection load (Watt) (R290) 437 | Dual comp. 781
Start current (R290) 13,1 A
Sound level 55.3 dB(A)
Air system BioLine ventilated air distribution system
Defrost system Automatic smart defrost with re-evaporation of defrost water

*   Dimensions in mm
**  100% operation at +23 °C (+/- 1K) ambient temperature

EF660W

Temperature range -35 °C/-5 °C
Ambient temperature range +10 °C/+30 °C
Control unit Gram Control Unit with voltage free contact, E-sensor and offset function
Alarms Acoustic and visual, High/Low temperature alarms and door alarm
Access port 1 pcs. Ø 24.5 mm
Material interior Stainless steel and inner doors in PMMA
Material exterior White lacquered steel or stainless steel
Dimensions (W x D x H) 837 x 756 x 2025/2275*
Net weight Unpackaged net weight without optional fittings: 152 kg
Gross volume 660 litres
Net volume 484 litres
Modules for shelves and drawers Shelves 27 - Drawers 13
Insulation 60 mm polyurethane with HFC-free cyclopentane propellant
K-Value Solid door 0.31 W/m²K
Refrigerant R290
Refrigerant charge 92 g
Refrigeration capacity at -40°C R290 244 watt
CO2e -
Energy consumption (kWh/24h) (R290) 9.02 | Dual comp. 12.87
Heat emission (Watt) (R290) 447,17 | Dual comp. 838,93**
Connection 230 V, 50 Hz
Connection load (Watt) (R290) 437 | Dual comp. 781
Start current (R290) 19,5 A
Sound level 55.3 dB(A)
Air system BioLine ventilated air distribution system
Defrost system Automatic smart defrost with re-evaporation of defrost water

*   Dimensions in mm
** 100% operation at +23°C (+/- 1K) ambient temperature

BioPlus 500

All the benefits of the BioPlus range, on a small footprint

BioPlus 500
BioPlus 600/660D

A deep cabinet, with good proportions and performance

BioPlus 600/660D
BioPlus 600/660W

A wide cabinet, enabling superior overview of the items

BioPlus 600/660W
BioPlus 930

Largest single door BioLine cabinet

BioPlus 930
BioPlus 1270/1400

Double door models with exceptional capacity

BioPlus 1270/1400

Choosing the right product..

“Application” icons bridge the gap between product ranges and the requirements in the different segments, work environments and use patterns. Allowing for quick and precise evaluation of needs and requirements.

Customised to your needs

Every single user has a unique set of requirements for how a biostorage fridge or freezer should be configured, and how the interior should be laid out. That’s why there are no standard arrangements for cabinets in the BioLine range – it’d be highly unlikely that “standard” would live up to your particular working requirements.

Our approach is to make sure that all the BioLine cabinets featured have a high standard specification. 

We then provide you with a comprehensive selection of optional features, as well as standardised modules for internal arrangements, fittings and accessories. 
These choices enable you to select exactly what you need, and to customise both the operating specifications and the interior layout of your BioLine cabinets, so that they’re right in line with your particular requirements.

With a cabinets service life expected to be many years, we argue that using a little more time in selecting the configuration that suites your exact specifications, is worth it.

All BioLine cabinets have a flexible layout, and can be altered on a whim; and if the need should arise, you can easily acquire new fittings according to the changing needs.