BioCompact II 410

ATEX-marking
RR410
II 3G Ex ec nC ic llB T6 Gc (R600a)
DTI 22ATEX0251X

ATEX-marking
RF410
II 3G Ex ec nC ic llB T3 Gc
DTI 22ATEX0251X

The BioCompact II 410 is a large refrigerator or freezer cabinet that offers a balanced blend of storage capacity, performance, and flexibility of a very small footprint.

Using a finned tube evaporator and forced air distribution is unique for a biostorage cabinet in this segment. The design provides more refrigerated storage space without cold walls that could potentially damage any delicate items stored in the cabinet. The design also provides additional key features such as a flexible interior layout, automatic re-evaporation of defrost water as well as internal and external ATEX compliance.

The BioCompact II 410 can be fitted to your needs as a refrigerator RR (+2°C/+20°C) or freezer RF (-25°C/-5°C) and is available with an ABS interior and white or stainless-steel exterior finish.

Scroll down to the accessory panel to see what fittings you can equip the cabinet with.

RR410

Temperature range +2 °C/+20 °C
Ambient temperature range Solid door +10 °C/+35 °C; glass door +10 °C/+32 °C
Control unit Gram Control Unit with voltage free contact, alarms, and offset function
Alarms Acoustic and visual, High/Low temperature alarms and door alarm
Access port 1 pcs. Ø 24.5 mm
Material interior PS lining
Material exterior White lacquered steel or stainless steel finish
Dimensions (W x D x H) 595 x 640 x 1776/1976*
Net weight Unpackaged net weight without optional fittings: 78 kg
Gross volume 346 litres
Net volume 312 litres
Modules for shelves and drawers 25 shelves, 13 drawers and 8 wire baskets
Insulation 50 mm polyurethane with HFC-free cyclopentane propellant
K-Value Solid door 0.35 W/m²K; glass door 0.51 W/m²K
Refrigerant R600a
Refrigerant charge 60 g
Refrigeration capacity at -10°C R600a 207 watt
CO2e -
Energy consumption (kWh/24h) (R600a) Solid door: 0.59 / Glass door: 1.31
Heat emission (Watt) (R600a) Solid door: 109,58 / Glass door: 164,69**
Connection 230 V, 50 Hz
Connection load (Watt) (R600a) Solid door: 113 / Glass door: 177
Start current (R600a) Solid door: 6,4 A / Glass door: 8,6 A
Sound level 33,2 dB(A)
Air system BioLine ventilated air distribution system
Defrost system Automatic smart defrost with re-evaporation of defrost water

RF410

Temperature range -25 °C/-5 °C
Ambient temperature range +10 °C/+35 °C
Control unit Gram Control Unit with voltage free contact alarm and offset function
Alarms Acoustic and visual, High/Low temperature alarms and door alarm
Access port 1 pcs. Ø 24.5 mm
Material interior PS lining
Material exterior White lacquered steel or stainless steel
Dimensions (W x D x H) 595 x 640 x 1776/1976*
Net weight Unpackaged net weight without optional fittings: 78 kg
Gross volume 346 litres
Net volume 312 litres
Modules for shelves and drawers 25 shelves, 13 drawers and 8 wire baskets
Insulation 50 mm polyurethane with HFC-free cyclopentane propellant
K-Value Solid door 0.35 W/m²K
Refrigerant R600a
Refrigerant charge 47 g
Refrigeration capacity at -25°C R600a 204 watt
CO2e -
Energy consumption (kWh/24h) (R600a) 2.06
Heat emission (Watt) (R600a) 153,4**
Connection 230 V, 50 Hz
Connection load (Watt) (R600a) 157
Start current (R600a) 8 A
Sound level 36,6 dB(A)
Air system BioLine ventilated air distribution system
Defrost system Automatic smart defrost with re-evaporation of defrost water

*    Dimensions in mm
**  100% operation at +23 °C (+/- 1K) ambient temperature

BioCompact II 210

Under bench refrigerator or freezer, applying itself to a wide

BioCompact II 210
BioCompact II 310

Intermediate sized cabinet, versatile layout, suited to

BioCompact II 310
BioCompact II 610

Incorporating pivotal features from the top BioLine models

BioCompact II 610
BioCompact II 210/210 combination

Combination of 2 x 210 stacked, you chose the

BioCompact II 210/210 combination
BioCompact II RR310/RF210 combination

The BioCompact II 310/210 combination

BioCompact II RR310/RF210 combination

Choosing the right product..

“Application” icons bridge the gap between product ranges and the requirements in the different segments, work environments and use patterns. Allowing for quick and precise evaluation of needs and requirements.

Customised to your needs

Every single user has a unique set of requirements for how a biostorage fridge or freezer should be configured, and how the interior should be laid out. That’s why there are no standard arrangements for cabinets in the BioLine range – it’d be highly unlikely that “standard” would live up to your particular working requirements.

Our approach is to make sure that all the BioLine cabinets featured have a high standard specification. 

We then provide you with a comprehensive selection of optional features, as well as standardised modules for internal arrangements, fittings and accessories. 
These choices enable you to select exactly what you need, and to customise both the operating specifications and the interior layout of your BioLine cabinets, so that they’re right in line with your particular requirements.

With a cabinets service life expected to be many years, we argue that using a little more time in selecting the configuration that suites your exact specifications, is worth it.

All BioLine cabinets have a flexible layout, and can be altered on a whim; and if the need should arise, you can easily acquire new fittings according to the changing needs.